THESE NUMBERS ARE VALID AS OF 15Mar95 (tmc)

_____________________________________________________________

 

Here are the new values for the encoder and steps for 230Rx:

 

CONTROLLER AXIS DESCRIPTION ENCODER STEPS MACRO

---------------------------------------------------------------------

1 1 4th rotary 1047 -6100 EM21

1 2 5th rotary 42907 68650 EM22

1 3 Y-factor mirror To move IN type 3ma-34000 OUT type 3ma0

1 5 230/490 Rx foc 849 -14400 EM25

 

2 1 345/650 Rx foc -2368 19000 EM26

2 2 5th horizontal -1320 13025 EM27

2 3 5th vertical -299 3000 EM28

2 4 4th linear -3849 -15400 EM29

=====================================================================

 

 

and now the new values for the 490Rx:

 

CONTROLLER AXIS DESCRIPTION ENCODER STEPS MACRO

---------------------------------------------------------------------

1 1 4th rotary 1409 -8040 EM41

1 2 5th rotary 43406 69435 EM42

1 3 Y-factor mirror To move IN type 3ma-35500 OUT 3ma0

1 5 230/492 Rx foc 472 -3800 EM45

 

2 1 345/650 Rx foc -2368 19000 EM46

2 2 5th horizontal 2670 -26700 EM47

2 3 5th vertical -565 5650 EM48

2 4 4th linear -204 -800 EM49

=====================================================================

 

 

and now the new values for the 345Rx:

 

CONTROLLER AXIS DESCRIPTION ENCODER STEPS MACRO

---------------------------------------------------------------------

1 1 4th rotary 1538 -8900 EM31

1 2 5th rotary -136889 -219015 EM32

1 3 Y-factor mirror To move IN type 3ma-35500 OUT 3ma0

1 5 230/490 Rx foc 2105 -17000 EM35

 

2 1 345/650 focus -2021 16200 EM36

2 2 5th horizontal 3082 -30820 EM37

2 3 5th vertical -586 5885 EM38

2 4 4th linear 875 3500 EM39

====================================================================

 

and now the new values for the 650Rx:

 

CONTROLLER AXIS DESCRIPTION ENCODER STEPS MACRO

---------------------------------------------------------------------

1 1 4th rotary 1430 -8170 EM61

1 2 5th rotary -137253 -218300 EM62

1 3 Y-factor mirror To move IN type 3ma-35500 OUT 3ma0

1 5 230/490 Rx foc 2105 -17000 EM65

 

2 1 345/650 focus -2616 21000 EM66

2 2 5th horizontal -1121 11200 EM67

2 3 5th vertical -586 5885 EM68

2 4 4th linear -2031 -8110 EM69

=====================================================================

USE THE FOLLOWING INSTRUCTIONS TO MOVE TO THE 230 RECEIVER:

____________________________________________________________

Turn on the Toshiba 3200 computer and when you get the DOS prompt (C:\>)

 

type MOTOR <enter>

 

This will start the Procomm communications program.

Connect the serial cable from SMC 2.

 

To move the 5th mirror horizontal motion:

 

type em27 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em27 again.

To move the 5th mirror vertical motion:

 

type em28 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em28 again.

To move the 4th mirror linear motion:

 

type em29 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em29 again

 

Now disconnect the serial cable from SMC 2 and connect the cable from SMC 1.

 

To move the 4th mirror rotary motion:

 

type em21 <enter>

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em21 again

To move the 5th mirror rotary motion:

 

type em22 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em22 again

 

Now make sure the 230/490 receiver is in the correct focus position:

 

type em25 <enter>

 

There may not be any movement at this point because the receiver focus for

the 230/490 Rx does not normally change.

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em25 again

 

type <alt>X and then type Y

to exit the Procomm program and then turn off the Toshiba 3200 computer,

disconnect the serial cable, and close the display so the receiver plate

will not hit it during a ZA slew.

============================================================================

USE THE FOLLOWING INSTRUCTIONS TO MOVE TO THE 490 RECEIVER:

____________________________________________________________

Turn on the Toshiba 3200 computer and when you get the DOS prompt (C:\>)

 

type MOTOR <enter>

 

This will start the Procomm communications program.

Connect the serial cable from SMC 2.

 

To move the 5th mirror horizontal motion:

 

type em47 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em47 again

To move the 5th mirror vertical motion:

 

type em48 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em48 again

To move the 4th mirror linear motion:

 

type em49 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em49 again

 

Now disconnect the serial cable from SMC 2 and connect the cable from SMC 1.

 

To move the 4th mirror rotary motion:

 

type em41 <enter>

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em41 again

To move the 5th mirror rotary motion:

 

type em42 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em42 again

 

Now make sure the 230/490 receiver is in the correct focus position:

 

type em45 <enter>

 

There may not be any movement at this point because the receiver focus for

the 230/490 Rx does not normally change.

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em45 again

 

type <alt>X and then type Y

to exit the Procomm program and then turn off the Toshiba 3200 computer,

disconnect the serial cable, and close the display so the receiver plate

will not hit it during a ZA slew

============================================================================

USE THE FOLLOWING INSTRUCTIONS TO MOVE TO THE 345 RECEIVER:

____________________________________________________________

Turn on the Toshiba 3200 computer and when you get the DOS prompt (C:\>)

 

type MOTOR <enter>

 

This will start the Procomm communications program.

Connect the cable from SMC 1.

 

To move the 4th mirror rotary motion:

 

type em31 <enter>

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em31 again

To move the 5th mirror rotary motion:

 

type em32 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em32 again

 

Disconnect the serial cable from SMC 1 and connect the cable from SMC 2.

 

type em36 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em36 again.

 

This will move the 345/650 receiver to the correct focus position.

 

To move the 5th mirror horizontal motion:

 

type em37 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em37 again

To move the 5th mirror vertical motion:

 

type em38 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em38 again

To move the 4th mirror linear motion:

 

type em39 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em39 again

 

You are now finished moving the mirrors and receiver focus.

 

type <alt>X and then type Y

to exit the Procomm program and then turn off the Toshiba 3200 computer,

disconnect any serial cable, and close the display so the receiver plate

will not hit it during a ZA slew.

============================================================================

USE THE FOLLOWING INSTRUCTIONS TO MOVE TO THE 650 RECEIVER:

____________________________________________________________

Turn on the Toshiba 3200 computer and when you get the DOS prompt (C:\>)

 

type MOTOR <enter>

 

This will start the Procomm communications program.

Connect the cable from SMC 1.

 

To move the 4th mirror rotary motion:

 

type em61 <enter>

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em61 again

To move the 5th mirror rotary motion:

 

type em62 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em62 again

 

Disconnect the serial cable from SMC 1 and connect the cable from SMC 2.

 

type em66 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em66 again.

 

This will move the 345/650 receiver to the correct focus position.

 

To move the 5th mirror horizontal motion:

 

type em67 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em67 again

To move the 5th mirror vertical motion:

 

type em68 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em68 again

To move the 4th mirror linear motion:

 

type em69 <enter>

 

The display will show "0 nnnn" the final encoder value.

The number shown should agree within about +/-3 of the table values.

If they do not agree close enough then type em69 again

 

You are now finished moving the mirrors and receiver focus.

 

type <alt>X and then type Y

to exit the Procomm program and then turn off the Toshiba 3200 computer,

disconnect any serial cable, and close the display so the receiver plate

will not hit it during a ZA slew.